News Technical Data Downloads

Main unit

Net weight: 0.95 kg
Measurements (LxWxH) 21 cm x 12cm x 8 cm
Batteries: 3 x Li-Ion 3.6 V, 1.5 Ah rechargeable
Power consumption: 1 W stand by (laser diode off)
20 W during laser operation
Infrared Laser-diode:
Wavelength: 980 nm
Operation modes: continuous wave mode (cw)
pulse mode
Power capability of laser diode: 7.5 +/- 1.5 W cw and peak power
Output power ex fiber: 0.5 - 5 W cw adjustable range, 20 % tolerance
3.75 +/-0.75 W time average in pulse mode 50% duty cycle
Pulse mode frequency range: 1 Hz – 300 Hz, 1% tolerance
Pulse duration range: 1 ms – 1 s, tolerance 0.2 ms
Intermittent operation:

5 min ON / 10 min OFF

Visible Laser-diode:
Wavelength: 660 nm
Operation modes: continuous wave mode (cw)
pulse mode
Power capability of laser diode: 100+/-20 mW cw/qcw
Output power ex fiber: 10 - 100 mW cw adjustable range, 20 % tolerance
50+/-10 mW time average in pulse mode 50% duty cycle
Pulse mode frequency range: 1 Hz – 300 Hz, 1% tolerance
Pulse duration range: 1 ms – 1 s, tolerance 0.2 ms
Intermittent operation: 10 min ON / 10 min OFF
Fiber optic receptacle:
Plug type: F-SMA 905 (3.16 mm ferrule), high power variant
Fiber core diameter: ≥ 200 um
Beam divergence at fiber exit NA=0.22 +/- 0.2
(total included angle of cone is 25°+/-2.5°)
Charger (supply unit)
Net weight: 0.25 kg
Measurements (L x W x H): 15 cm x 10 cm x 5 cm
Mains supply voltage input: 100 – 240 V AC, 50 – 60 Hz
Input supply current: 1 A
Output voltage: 5.7 V DC
Output current: 0 – 4 A
Output voltage ripple: 0.5 V eff.
Power consumption: 0.6 W during stand by
28 W with nomimal load current
Foot switch (radio link remote control unit)
Net weight: 0.1 kg
Measurements (LxWxH): 10 cm x 7 cm x 2 cm
Battery: Li-Ion prismatic, 3.6 V, 0.5 Ah, rechargeable
Power consumption: 0.2 W
Frequency band: 2.45 GHz
Transmission power: 2.4 mW
Transmission standard: IEEE 802.15.4
Operation conditions
ambient temperature range: 15° C to 30° C, (must be above dew point)
relative humidity range: 30% to 75% (not condensing)
atmospheric pressure range: 865 hPa to 1085 hPa
Storage and transport conditions
ambient temperature range: -10° C to 40° C, (must be above dew point)
relative humidity range: 30% to 80% (not condensing)
atmospheric pressure range: 300 hPa to 1085 hPa