Main unit |
 |
 |
|
| Net weight: |
0.95 kg |
 |
|
| Measurements (LxWxH) |
21 cm x 12cm x 8 cm |
 |
|
| Batteries: |
3 x Li-Ion 3.6 V, 1.5 Ah rechargeable |
 |
|
| Power consumption: |
1 W stand by (laser diode off)
20 W during laser operation |
|
 |
| Infrared Laser-diode: |
 |
|
|
 |
Visible Laser-diode:
|
 |
 |
|
| Wavelength: |
660 nm |
 |
|
| Operation modes: |
continuous wave mode (cw)
pulse mode |
 |
|
| Power capability of laser diode: |
100+/-20 mW cw/qcw |
 |
|
| Output power ex fiber: |
10 - 100 mW cw adjustable range, 20 % tolerance
50+/-10 mW time average in pulse mode 50% duty cycle
|
 |
|
| Pulse mode frequency range: |
1 Hz – 300 Hz, 1% tolerance |
 |
|
|
|
| Pulse duration range: |
1 ms – 1 s, tolerance 0.2 ms |
|
|
| Intermittent operation: |
10 min ON / 10 min OFF |
|
 |
Fiber optic receptacle:
|
 |
 |
|
| Plug type: |
F-SMA 905 (3.16 mm ferrule), high power variant |
 |
|
| Fiber core diameter: |
≥ 200 um |
 |
|
| Beam divergence at fiber exit |
NA=0.22 +/- 0.2
(total included angle of cone is 25°+/-2.5°) |
|
 |
| Charger (supply unit) |
|
|
 |
|
| Net weight: |
0.25 kg |
 |
|
| Measurements (L x W x H): |
15 cm x 10 cm x 5 cm |
 |
|
| Mains supply voltage input: |
100 – 240 V AC, 50 – 60 Hz |
 |
|
| Input supply current: |
1 A |
 |
|
| Output voltage: |
5.7 V DC |
 |
|
| Output current: |
0 – 4 A |
 |
|
| Output voltage ripple: |
0.5 V eff. |
 |
|
| Power consumption: |
0.6 W during stand by
28 W with nomimal load current |
|
 |
| Foot switch (radio link remote control unit) |
 |
 |
|
| Net weight: |
0.1 kg |
 |
|
| Measurements (LxWxH): |
10 cm x 7 cm x 2 cm |
 |
|
| Battery: |
Li-Ion prismatic, 3.6 V, 0.5 Ah, rechargeable
|
 |
|
| Power consumption: |
0.2 W |
 |
|
| Frequency band: |
2.45 GHz |
 |
|
| Transmission power: |
2.4 mW |
 |
|
| Transmission standard: |
IEEE 802.15.4
|
|
 |
| Operation conditions |
 |
 |
|
| ambient temperature range: |
15° C to 30° C, (must be above dew point) |
 |
|
| relative humidity range: |
30% to 75% (not condensing) |
 |
|
| atmospheric pressure range: |
865 hPa to 1085 hPa |
|
 |
| Storage and transport conditions |
 |
 |
|
| ambient temperature range: |
-10° C to 40° C, (must be above dew point) |
 |
|
| relative humidity range: |
30% to 80% (not condensing) |
 |
|
| atmospheric pressure range: |
300 hPa to 1085 hPa |
|